Kraj: Australia
Język: angielski
Źródło: Department of Health (Therapeutic Goods Administration)
Medilink Pty Limited
Class I
KUB Technologies Inc 111 Research Drive, Stratford, CT, 6615 United States Of America
42280 - Mobile specimen x-ray system, digital
Cabinet x-ray systems verified to 0.1 milliroentgen at any point 5cm outside the external surface whereas US FDA 21CFR 1020.40 allows - Emission limit. (i) Radiation emitted from the cabinet x-ray system shall not exceed an exposure of 0.5 milliroentgen in one hour at any point five centimeters outside the external surface
A
2019-05-13