TERUFLEX BLOOD BAG SYSTEM ANTICOAGULANT CITRATE PHOSPHATE DEXTROSE (CPD) AND OPTISOL RED CELL PRESERVATIVE (anticoagulant citrat

מדינה: ארצות הברית

שפה: אנגלית

מקור: NLM (National Library of Medicine)

קנה את זה

הורד מאפייני מוצר (SPC)
11-12-2019

מרכיב פעיל:

Trisodium Citrate Dihydrate (UNII: B22547B95K) (Anhydrous Citric Acid - UNII:XF417D3PSL), SODIUM PHOSPHATE, MONOBASIC, UNSPECIFIED FORM (UNII: 3980JIH2SW) (Phosphate Ion - UNII:NK08V8K8HR, Sodium Cation - UNII:LYR4M0NH37), Dextrose Monohydrate (UNII: LX22YL083G) (Anhydrous Dextrose - UNII:5SL0G7R0OK), Anhydrous Citric Acid (UNII: XF417D3PSL) (Anhydrous Citric Acid - UNII:XF417D3PSL)

זמין מ:

Terumo Corporation

INN (שם בינלאומי):

Trisodium Citrate Dihydrate

הרכב:

Trisodium Citrate Dihydrate 26.3 g in 1000 mL

מסלול נתינה (של תרופות):

INTRAVENOUS

סוג מרשם:

PRESCRIPTION DRUG

סממני תרפויטית:

Revised 8/96 TERUFLEX® BLOOD BAG SYSTEM CPD/OPTISOL® SOLUTION Read these instructions before use. INSTRUCTIONS FOR BLOOD COLLECTION: Use aseptic technique 1. Confirm that all numbered tubing of each blood bag unit contains segment numbers. 2. Make a loose knot in the donor tubing approximately 10 cm from needle unless alternate methods are used to seal tubing. 3. Clamp donor tubing. 4. Suspend primary bag as far as possible below the donor's arm. 5. Apply blood pressure cuff or tourniquiet to donor's arm. Disinfect site of phlebotomy. If blood pressure cuff is used, inflate cuff to 60 mmHg. 6. Remove needle protector and perform phlebotomy. Remove clamp to permit blood flow into primary bag. CAUTION Do not touch needle after removing the needle protector. 7. Appropriately secure donor tubing to donor's arm. 8. MIX BLOOD WITH ANTICOAGULANT AT SEVERAL INTERVALS DURING COLLECTION. 9. Collect labeled volume of blood (+/- 10%). 10. Tighten knot firmly after collection. Clamp betwe

מצב אישור:

New Drug Application

מאפייני מוצר

                                TERUFLEX BLOOD BAG SYSTEM ANTICOAGULANT CITRATE PHOSPHATE
DEXTROSE (CPD) AND OPTISOL RED CELL PRESERVATIVE - ANTICOAGULANT
CITRATE
PHOSPHATE DEXTROSE (CPD) AND AS-5 RED CELL PRESERVATIVE
TERUMO CORPORATION
----------
TERUFLEX® BLOOD BAG SYSTEM CPD/OPTISOL® SOLUTION BLOOD BAG SYSTEM
Revised 8/96
TERUFLEX® BLOOD BAG SYSTEM
CPD/OPTISOL® SOLUTION
READ THESE INSTRUCTIONS BEFORE USE.
INSTRUCTIONS FOR BLOOD COLLECTION: USE ASEPTIC TECHNIQUE
1. Confirm that all numbered tubing of each blood bag unit contains
segment numbers.
2. Make a loose knot in the donor tubing approximately 10 cm from
needle unless alternate methods are
used to seal tubing.
3. Clamp donor tubing.
4. Suspend primary bag as far as possible below the donor's arm.
5. Apply blood pressure cuff or tourniquiet to donor's arm. Disinfect
site of phlebotomy. If blood
pressure cuff is used, inflate cuff to 60 mmHg.
6. Remove needle protector and perform phlebotomy. Remove clamp to
permit blood flow into primary
bag.
CAUTION Do not touch needle after removing the needle protector.
7. Appropriately secure donor tubing to donor's arm.
8. MIX BLOOD WITH ANTICOAGULANT AT SEVERAL INTERVALS DURING
COLLECTION.
9. Collect labeled volume of blood (+/- 10%).
10. Tighten knot firmly after collection. Clamp between knot and
needle. Sever donor tubing between
knot and clamp. Collect blood samples.
11. Reapply clamp to donor tubing; release pressure on donor's arm and
remove needle. Seal donor
tubing.
CAUTION Discard tubing/phlebotomy needle unit according to
institutional procedures.
12. Immediately after collection, invert bag several times to assure
blood and anticoagulant are well
mixed.
13. Strip blood from donor tubing into bag, mix well, and allow tubing
to refill. Seal on or near X marks
on donor tubing to provide numbered aliquots of anticoagulated blood
for testing.
14. Centrifuge unit to separate red cells from plasma.
15.The time of addition of OPTISOL Solution may vary depending on the
processing option selected.
Add solution under one of the following condi
                                
                                קרא את המסמך השלם